TRANSIENT3000
86/162
6.2.3.4 DIPS
loaded test EUT power v, i
possible ramps at DIP
Example Level Ramp: Amplitude as % of EUT voltage
loaded test EUT power v, i
Level ramp at DIP
-400
-300
-200
-100
0
100
200
300
400
6.2.3.4.1 Definition DIP level ramp
The start of the ramp begins at 100% and changes with „Level step“ (10%) to the DIP voltage. If the test
time has been selected to be long enough, the voltage UT will be reached.
Ramp: Length of interruption in ms:
loaded test EUT power v, i
Duration ramp at DIP
1 dip
Level step