Fault removal multi N/C 2100S
102
Control lamps at the analyzer do not illuminate: 5 V, 24 V
Cause Remedy
Error in the power supply or in the electronics check the electrical connections
check the power supply of the lab
equipment fuse faulty inform Service
Front LED display on the analyzer does not illuminate: Login
Cause Remedy
internal program has not been started switch analyzer on again (switch off/on from main switch)
Control lamp at analyzer does not illuminate: Heating
Cause Remedy
incorrect temperature configuration in multiWin check temperature configuration in multiWin under
ONFIGURATION EDIT OPTIONS on the tab ANALYZER
COMPONENTS
(list field FURNACE TEMPERATURE)
faulty thermocouple (furnace)
Note: A faulty thermocouple can be detected by an
indication in the LED strip in the analyzer
inform Service
faulty electronics component inform Service
combustion furnace not connected correctly check correct contact of the combustion furnace
Automatic lock leaking
Cause Remedy
automatic lock does not close correctly Open lock manually (see p. 18 ff.)
Apply manual metering syringe with septum, press slightly
and check flow (window SYSTEM STATE)
Check the gas pressure of the auxiliary gas for the locks
(4 to 6 bar preliminary pressure required)
Check the hose connections of the lock
Replace the septum in the lock
if necessary inform Service