multi N/C 2100S Maintenance and care
63
7 Maintenance and care
7.1 Maintenance intervals
Analyzer
Maintenance task Maintenance interval
Clean and maintain the device weekly
Clean drip tray and reagent bottle weekly and after filling
Inspect all hose connections for tight fit monthly
Inspect the fastening screws for tight fit monthly
Water traps
Maintenance task Maintenance interval
Check gas flow daily
Exchange water trap as required, but no later than after 6 months
Halogen trap
Maintenance task Maintenance interval
Check copper wool for discoloring daily
Replace depleted copper/brass wool if half of the copper wool or the brass wool is
discolored
Combustion tube
Maintenance task Maintenance interval
Check for cracks and damages during catalyst replacement
Check catalyst and replace if necessary as required, after corresponding message in
multiWin at the latest
Clean combustion tube during catalyst replacement
Renew (replace) combustion tube
Recommendation:
replace combustion tube and catalyst
as required, but no later than after 12
months
TIC condensate container
Maintenance task Maintenance interval
Check for cracks and damages 3 months
Cleaning the TC condensation vessel as required, but no later than after 12
months
Condensation coil
Maintenance task Maintenance interval
Check for cracks and damages 3 months
Clean the condensation coil as required, but no later than after 12
months