Operation multi N/C 2100S
62
3. In the window S
YSTEM STATE select sample supply with sampler by clicking the button
[S
AMPLER].
4. Fill the sample cups with the measuring liquid and place them onto the sample tray.
5. Only for NPOC measurements:
Fill the acid cup with HCl (c = 2 mol/l) and place the cup into the acid position of the
sample tray.
6. Start the measurement.
Click on [S
TART MEASUREMENT] or open the menu command MEASUREMENT START
MEASUREMENT.
The window M
EASUREMENT START opens.
In the window MEASUREMENT START
enter a name for a new analysis table or select an
existing analysis table with [E
DIT].
With [START ] open the window CURRENT SAMPLE DATA.
Open an existing rack table or enter the sample name in the column S
AMPLE ID in
accordance with the assignment of the sample rack. You can also enter the dilution,
sample type and unit.
With [] release the samples.
Confirm the entries with [].
The rack table will be closed.
A query follows whether the rack table should be saved. If you want to reuse the
entries later, open the default window for saving files with [Y
ES].
Next the window M
EASUREMENT opens. Start the measurement with [START F2] and
follow the instructions of the control and analysis software.
At the end of the measurement the results appear in the selected analysis table.