Fault removal multi N/C 2100S
92
8.2 Error messages in multiWin
Error code Error message
VERS Communication error - incorrect command set between PC and device!
Cause Remedy
the internal and external program versions
do not match
update the internal and external program
VERS1 Communication error - analyzer
Cause Remedy
analyzer not switched on switch on analyzer
multiWin started too early only start multiWin after 30 sec
analyzer not connected to PC check connection between analyzer and PC
incorrect COM port set in multiWin at the
external computer
check set interface at the external computer, if
necessary select different interface in multiWin with
menu command Configuration / interface
-6 Analysis device is busy
Cause Remedy
analyzer > 10 min in the busy state initialize analyzer
-5 communication error - analyzer STAT, MESS, STEP or INIT
Cause Remedy
Communication errors initialize analyzer
-4 Communication error - analyzer
Cause Remedy
Communication errors check interface cable
initialize analyzer
-3
-2
-1
command from the analyzer CRC error CRC error
invalid command from the analyzer
Cause Remedy
Communication errors initialize analyzer