Operation multi N/C 2100S
60
6.4 Performing measurements
Note for multiWin pharma: Only methods with the signatures status "authorized" can be
used for the measurement.
6.4.1 Measurement with manual sample supply
NOTICE
For each injection a maximum of 500 μl measurement substance may be injected. Inject
the sample only after the prompt by the control and evaluation software and in the
stated intervals!
Always ensure that the volume set in the method is injected!
Carry out a measurement with manual sample supply as follows:
1. In the window S
YSTEM STATE check the following entries:
Visual bank OK (not for multi N/C 2100S pharma)
if applicable CLD or, if applicable, ChD - OK
Gas flow - OK
Temperature - OK
Note:
If one of the entries is incorrect (shown in red), carry out a fault finding exercise in
accordance with the notes in section "Fault removal" p. 91.
2. In the window S
YSTEM STATE select the manual sample supply by clicking the button
[
MANUAL].
3. Prepare the analyzer for the measurement with manual sample supply as follows:
Slide the septum supplied onto the syringe cannula up to the union nut.
Note:
The septum guarantees the system tightness for the septum-free lock during injection.
4. Using the menu command M
ETHOD NEW create a new method or load an existing
method. To do so open via the menu command M
ETHOD LOAD the database window
M
ETHOD SELECTION, highlight the desired method and confirm the selection by clicking
the button [OK].
This is followed by the initialization of the analyzer.
5. Start the measurement.
Click on [S
TART MEASUREMENT] or open the menu command MEASUREMENT START
MEASUREMENT.
The window M
EASUREMENT START opens.
Enter the sample ID and, if applicable, a name for the analysis table. You can
also enter the dilution, sample type and unit.
With [S
TART ] open the window MEASUREMENT.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Endress+Hauser Analytik Jena multi N/C 2100S and is the answer not in the manual?

Endress+Hauser Analytik Jena multi N/C 2100S Specifications

General IconGeneral
BrandEndress+Hauser
ModelAnalytik Jena multi N/C 2100S
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals