multi N/C 2100S Maintenance and care
77
3. Slide the three coated sealing rings (3)
onto the combustion tube.
Note:
Make sure that the sealing rings at the edge
of the combustion tube are flush.
4. Position the ceramic holder in the top
opening of the combustion furnace.
5.
Insert the combustion tube into the
combustion furnace.
6. Screw the lock with the Allen screw
loosely into the holder.
7.
Hold the combustion tube in place from
below. Carefully place the TC lock onto
the combustion tube up to the stop;
press the lock slightly against it and
tighten the union nut
FINGER-TIGHT.
8. Screw the lock with the Allen screw
loosely into the holder.
9. Screw the Fingertight connection of the
carrier gas connection into the TC lock.
10. Attach the bottom end of the
combustion tube and the inlet of the
condensation coil (spherical joint (3)).
11.
Secure the spherical joint with the fork
clamp (2) and tighten the knurled head
screw (1) finger-tight.