Maintenance and care multi N/C 2100S
82
4. Press the condensation coil into the
clamps on the carrier plate (arrow.
Connect the following hoses:
• Hose no. 1 connects the TIC
condensate container to the cooling
coil
• Hose no. 16 to the condensate pump
• Hose no. 21 to the phosphoric acid
pump (this hose must protrude into
the TIC condensate container)
The FAST connector and hose no. 6 must be
pushed at least 1 cm onto the glass port of
the TIC condensate container.
6. Hang the carrier plate into the mount on
the right-hand side of the furnace.
The spherical joint of the condensation
coil will reach into the cut-out of the
combustion furnace.
Connect the Peltier cooling block at the
backplate (see arrow).
8. Connect hose no. 2 (to the water traps) to
the top outlet of the TIC condensate
container.