multi N/C 2100S Fault removal
99
Incorrect injection volume The sample volume set in the method has to be injected
(manual metering!)
faulty metering check metering (see p. 100)
particulate samples not or insufficiently stirred stir particulate samples
z coordinate adjustment for automatic locks not ok
(system leaking during injection into automatic lock)
check sampler adjustment and readjust sampler if
necessary (see p. 65)
For automatic locks: septum on syringe is no longer
tight
replace septum on syringe
Low results for analyses through combustion (TC, TOC, NPOC, TN
b
)
TIC measurements are ok
Cause Remedy
Catalyst depleted when using platinum catalysts for multi N/C and
measuring in the differential mode (neutral samples,
slightly alkaline) the catalyst can be regenerated
multiple injection (6 times) of acidified ultrapure water
(with HCl to pH 2) or catalyst replacement if regeneration
unsuccessful or not lasting
it is recommended to use 1 - 2 sample cups with acidified
ultrapure water for regeneration for each sample tray
for measurements in NPOC mode (acidic samples) replace
the catalyst
when using special catalyst for multi N/C replace the
catalyst
Note: After a catalyst replacement a calibration must be
carried out.
Use a cannula suitable for particles if the substance
contains particles
particulate samples not or insufficiently stirred stir particulate samples
u coordinate adjustment for automatic TC lock
incorrect
check sampler adjustment and readjust sampler if
necessary (see p. 65)
Low results for TIC measurements
Analyses through combustion (TC, TOC, NPOC) are ok
Cause Remedy
no phosphoric acid in the reagent bottle for phosphoric
acid
refill phosphoric acid
faulty metering of the sample check metering (see below)
septum faulty replace septum
Low results for TN
b
Cause Remedy
Catalyst depleted Exchanging the catalyst
measurement outside the calibrated range Pay attention to calibrated range
use quadratic calibration or multiple range calibration
Calibrate dependent on matrix when possible

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Endress+Hauser Analytik Jena multi N/C 2100S and is the answer not in the manual?

Endress+Hauser Analytik Jena multi N/C 2100S Specifications

General IconGeneral
BrandEndress+Hauser
ModelAnalytik Jena multi N/C 2100S
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals