OXY5500 Optical Oxygen Analyzer BA02195C
5
14 Endress+Hauser
Measurement values
Table 5–7 Definition of register 4109
Start
Register
Number of
Registers
Reg3 / Reg4
Write
Access
4109 2 Int32: Device ID of RS-
232. Minimum 1, Maxi-
mum 32.
Yes
Table 5–8 Definition of register 5695
Start
Register
Number of
Registers
Reg3 / Reg4
Write
Access
5695 2 Int32: Device ID of
Ethernet. Minimum 1,
Maximum 32.
Yes
Table 5–9 Definition of register 4895
Start
Register
# of
Registers
Reg1/
Reg2
Reg3/
Reg4
Reg5/
Reg6
Reg7/
Reg8
Reg9/
Reg10
Reg11/
Reg12
Reg13/
Reg14
Write
Access
4895 14 Float:
Pres-
sure in
mbar
Float:
Refer-
ence
Ampli-
tude in
mV
Float:
Oxy-
gen
Ampli-
tude in
mV
Float:
Oxy-
gen
Phase
shift in
degrees
Float:
Tem-
pera-
ture in
° C
Float:
Calcu-
lated
Oxygen
Value in
the unit
defined
in regis-
ter 2089
Int32:
Error Reg-
ister. See
Table 5–10
No
It is not necessary to read out all 14 registers. For simple
applications, reading out registers 9 to 14 (starting with Register
4903) may be all that is required.