OXY5500 Optical Oxygen Analyzer BA02195C
4
52 Endress+Hauser
The Baudrate for the RS-485 channel can be set to 9600, 19200,
38400, 57600 or 115200.
The ID, which is used within the Modbus communication, can be set
to any value between 1 and 32.
Parity can be set to Even, Odd or None.
All settings are applied by pressing
.
TCP/IP settings
Use this screen to set TCP/IP. Refer to Figure 4–52.
Setting the Parity to “None” will also set the number of stop bits
to two. Odd and Even settings will use one stop bit.
Figure 4–51 Digitals - RS-485 settings