OXY5500 Optical Oxygen Analyzer BA02195C
4
26 Endress+Hauser
Device Settings Menu Options
Press the Device Settings button on the MAIN MENU to access the DEVICE
SETTINGS menu, the SENSOR DETAILS screen and the ABOUT screen.
Device settings screen
This screen is used to change the general settings of the OXY5500. Refer to
Figure 4–28. Date, Time, LED Intensity (User Signal Intensity) and Forced Zero
settings are saved with every measurement in the respective measurement
file.
Time: Set the current time in hour (h), minute (m) and seconds (s).
Date: Set the current date in day (d), month (m) and year (y).
LED Intensity/User Signal Intensity: Adjusts the signal strength of
the probe.
If the power supply to the analyzer is disabled, Time and Date
settings will be set to zero. Reset Time and Date before starting a
new measurement so that the correct time is stored with the data.
The OXY5500 uses 24-hour time settings.
The LED Intensity (also called User Signal Intensity) setting range
is from -5 to 5, with 5 as the highest probe intensity and -5 the
lowest probe intensity. The default value is 0.