BA02195C OXY5500 Optical Oxygen Analyzer
Endress+Hauser B
–7
Required tools
• Flathead screwdriver (standard size and mini)
• 9/16 in. open-ended wrench
• Adjustable wrench
• 10 in. crescent wrench
Removing the pressure sensor
1. Turn off the power to the analyzer and open the enclosure door using
a standard flathead screwdriver to unlatch the lock.
2. With a 9/16 in. wrench, loosen the Swagelok nut closest to the
pressure sensor.
3. Using the same wrench, loosen the Swagelok nut on the T-fitting.
Refer to Figure B–6.
Figure B–6 Removing the Swagelok nuts
PRESSURE
SENSOR NUT
T-FITTING NUT