BA02195C OXY5500 Optical Oxygen Analyzer
Endress+Hauser 4
17
“Error code: 5” = No RTD (Pt100) and amplitude too low (bit 0 [2
N
Value 1], bit 2 [2
N
Value 4)=5)
“Error code: 1024” = No pressure sensor connected (bit 10)
“Error code: 1029” = No RTD (Pt100), amplitude too low, no pressure
sensor connected (bit 0 [2
N
Value 1], bit 2 [2
N
Value 4], bit 10 [2
N
Value 1024] = 1029)
Graph screen
The oxygen values of the current measurement session are displayed in a
graph; the last measurement value of the Current Measurement is shown at
the top of the screen. Refer to Figure 4–20.
Table 4–1 Error codes
Bit
2
N
Value
Error
0 1 No RTD (Pt100)
1 2 No sensor selected
2 4 Amplitude too low
3 8 SD card defect
4 16 Reference amplitude out of range
5 32 Photo diode saturated
6 64 Signal overflow
7 128 Signal overflow
8 256 Reserved
9 512 Critical error. Refer to “Service” on page B-19.
10 1024 No pressure sensor / pressure sensor out of range
11 2048 Reserved
12 4096 Storage space full