OXY5500 Optical Oxygen Analyzer BA02195C
5
4 Endress+Hauser
Holding Registers
Refer to Table 5–1 for table register definitions. When reviewing the table, it is
important to remember:
The register addresses referred to in Table 5–1 show the first address
of multiple addresses available per register (see “Size” column for
number of addresses per register). Do not add or subtract “1” from
the first address register number as this could cause conflict with
other register assignments.
The analyzer does not check for correct ranges. The host must ensure
that valid numbers are used. Any wrong value may lead to
unexpected performance.
If there is a Service Software connected (via USB), the Modbus
write command 16 (“Write multiple registers”) will always return
error code 6.
Table 5–1 Holding registers
Register
Name
Address Size
Variable
Type
Description
Write
Access
Firmware
Date
1023 8 Character Firmware creation date,
e.g., “2014-11-18\0\0”
(November 18, 2014)
No
Firmware
Version
1031 8 Character Firmware version, e.g.,
,,SSI v1.0.1.0287\0”
No
Serial Num-
ber
1063 8 Character Serial number, e.g.,
,,SAAP0000000001\0\0”
No
Oxygen Unit 2089 2 Int32 The oxygen unit that dis-
plays on the LCD of the
analyzer and also in mea-
surement register 4909.
Yes
Compensa-
tion tempera-
ture
2411 2 Float Sets the compensation
temperature. Refer to
Table 5–4.
Yes
Interval rate 3499 2 Mixed Sets the oxygen measure-
ment interval rate and also
deactivates oxygen mea-
surement. Refer to
Table 5–5. Range:
1 to 359999 seconds
Yes