BA02195C OXY5500 Optical Oxygen Analyzer
Endress+Hauser 5
5
Table 5-1 Holding registers (Continued)
Register
Name
Address Size
Variable
Type
Description
Write
Access
Device ID
RS-485
4095 2 Int32 Sets the device ID that is
used in Modbus RTU com-
munication (range 1-32).
Refer to Table 5–6.
Yes
Device ID
Minimum
RS-485
4097 2 Int32 Device ID Address Limit:
Minimum.
No
Device ID
Maximum
RS-485
4099 2 Int32 Device ID Address Limit:
Maximum.
No
Baud rate
RS-485
4101 2 Int32 Code for Baud rate where:
3 = 9600
4 = 19200
5 = 38400
6 = 57600
7 = 115200
Yes
Baud rate
Minimum
RS-485
4103 2 Int32 Minimum code for baud
rate.
No
Baud rate
Maximum
RS-485
4105 2 Int32 Maximum code for baud
rate.
No
Parity
RS-485
4107 2 Int32 Parity for the RS-485 out-
put where:
0x00 = Even Parity
0x01 = Odd Parity
0x02 = No Parity
1
1. When using “No Parity,” an additional Stop-bit will be sent (2 Stop bits).
Yes
Device ID
RS-232
4109 2 Int32 Sets the device ID used in
the Modbus RTU communi-
cation (range 1 to 32).
Refer to Table 5–7.
Yes
Device ID
Minimum
RS-232
4111 2 Int32 Device ID Address Limit:
Minimum.
No
Device ID
Maximum
RS-232
4113 2 Int32 Device ID Address Limit:
Maximum.
No