OXY5500 Optical Oxygen Analyzer BA02195C
5
6 Endress+Hauser
2. When using “No Parity,” an additional Stop bit will be sent (2 Stop bits).
Table 5-1 Holding registers (Continued)
Register
Name
Address Size
Variable
Type
Description
Write
Access
Baud Rate
RS-232
4115 2 Int32 Code for baud rate where:
0x03 = 9600
0x04 = 19200
0x05 = 38400
0x06 = 57600
0x07 = 115200
Yes
Baud Rate
Minimum
RS-232
4117 2 Int32 Minimum code for baud
rate.
No
Baud Rate
Maximum
RS-232
4119 2 Int32 Maximum code for baud
rate.
No
Parity
RS-232
4121 2 Int32 Parity for the RS-232 out-
put where:
0x00 = Even Parity
0x01 = Odd Parity
0x02 = No Parity
2
Yes
4-20mA
Port1 Output
Interface
4359 2 Int32 Code for 4-20mA Port1
Output Mode where:
0x00 = Off
0x01 = Fixed
0x02 = Linear
0x04 = Bilinear
Yes
4-20mA
Port1 Output
Channel
4363 2 Int32 Code for 4-20mA Port1
Output Interface where:
0x01 = Oxygen
0x20 = Temperature
Yes
4-20mA
Port1 Low
Value
4377 2 Float The 4mA output value.
Refer to “Analog input
and output value
ranges” on page 5-16 for
details.
Yes
4-20mA
Port1 Mid
Value
4379 2 Float The 12mA output value,
which is used in bilinear
mode only. Refer to “Ana-
log input and output
value ranges” on page
5-16.
Yes
4-20mA
Port1 High
Value
4381 2 Float The 20mA output value.
Refer to “Analog input
and output value
ranges” on page 5-16.
Yes