OXY5500 Optical Oxygen Analyzer BA02195C
4
–34 Endress+Hauser
• Pressure:
– Select Auto
to measure the atmospheric pressure via the
4-20mA input.
– Select Manual
if there is no pressure sensor connected to the
analyzer.
– Type in the current atmospheric pressure value and the
respective unit (hPa, mbar, PSI, atm or torr).
Press to save your changes.
Press the Next
button at the top right of the screen followed by .
Setting calibration temperature
Use the following instructions to program the analyzer for the correct
calibration temperature. Refer to Figure 4–37.
Figure 4–36 Calibration settings screen