BA02195C OXY5500 Optical Oxygen Analyzer
Endress+Hauser 4
–23
recommends 35 days as a starting point for re-calibration or as the application
requires it. Refer to Table 4–2 below and “Calibrating the analyzer” on page
4-35.
Logging and data management
On the MEASUREMENT SETTINGS screen, use the navigation buttons to move
to the Logging box. Refer to Figure 4–25.
• Select Off if you do not want to store measurement data.
• Select On to store measurement data.
The screen will switch to the Measurement Browser automatically. A
list will display showing the Measurement file name, the number of
measurement Points stored in the respective file and the date the file
was Last Used. Refer to Figure 4–26.
Table 4–2 Interval sampling rate/calibration frequency
Sampling Rate Points
Calibration
Frequency (Days)
30 seconds 100,000 34.7
1 minute 100,000 69.4
1 hour 100,000 4,166
10 hours 100,000 41,666
In the Status bar, the symbol indicates Logging is turned
off as shown in Figure 4–25.
Figure 4–25 Measurement settings screen -
Logging