OXY5500 Optical Oxygen Analyzer BA02195C
B
14 Endress+Hauser
11. Loosen the connector nut on the probe at the SMA connector inside
the analyzer enclosure. Refer to Figure B–16.
12. Carefully pull the probe out through the conduit and discard.
Installing the new oxygen probe
1. Carefully remove the protective plunger from the end of the probe
(analyzer side), taking care not to touch the optical fiber tip. Refer
to Figure B–17.
2. Feed the new probe through the conduit with the SMA connector end
entering first.
OXYGEN PROBE AND
SMA CONNECTOR NUT
Figure B–16 Remove connector nut (analyzer side)
TEE UNION
Optical fiber tip
DO NOT TOUCH
Figure B–17 Preparing the new oxygen probe