OXY5500 Optical Oxygen Analyzer BA02195C
4
–46 Endress+Hauser
2. Press to store the calibration data for the selected sensor.
The display will switch to the measurement screen automatically.
Relative accuracy test audit (RATA)
RATA is accessible from the RATA button on the SENSOR / SENSOR OPTIONS
menu screen.
Setting RATA
1. Select RATA in the SENSOR OPTIONS window. Refer to Figure 4–46.
Figure 4–46 Sensor options screen