BA02195C OXY5500 Optical Oxygen Analyzer
Endress+Hauser 4
53
If DHCP is selected, the IP and Subnet Mask will be assigned by the
DHCP server and therefore are not editable.
If Static is selected, the IP and Subnet Mask must be entered
manually. Contact your local Network Administrator for assistance if
input data needs to be confirmed.
Port specifies the network port under which the Modbus application
takes place. The default value for most Modbus applications is 502.
The ID, which is used with the Modbus communication, can be set to
any value between 1 and 32.
All settings are applied by pressing .
Analog Output Settings (Analogues) Menu Options
From the MAIN MENU, press Analogues to access the 4-20mA INTERFACE
SETTINGS, 4-20mA VALUES, CONCENTRATION ALARM RELAY (LS2) and
4-20mA CALIBRATION screens.
4-20mA interface settings
The 4-20mA INTERFACE SETTINGS can be accessed through the ANALOGUES
menu. When accessed, the following screen displays (refer to Figure 4–53).
Figure 4–52 Digitals - TCP/IP settings