BA02195C OXY5500 Optical Oxygen Analyzer
Endress+Hauser 4
–43
3. Select Yes to stop the measurement in order to move to the
CALIBRATION windows. Use the buttons to navigate
between input fields.
To enter editing mode
1. Press .
a. Change the setting or value (one digit at a time) by pressing the
buttons.
b. Make the desired change to an input field.
2. Press again to save the changes.
To exit editing mode
1. Press to cancel and exit.
Performing pre-calibration
1. Connect the analyzer to a nitrogen (N
2
) bottle.
2. Set the flow to 1.5 SLPM.
Figure 4–44 Message window - Stop
measurements during configuration