OXY5500 Optical Oxygen Analyzer BA02195C
B
6 Endress+Hauser
8. Remove the ground cable from the module.
a. Using the Philips screwdriver, remove the screw and cable.
Refer to Figure B–5.
9. Remove the electro-optical module from the enclosure and set aside.
Replacing the electro-optical module
1. Connect the ground cable to the replacement module.
2. Position the electro-optical module above the DIN rail and snap down
into place.
3. Re-wire the terminal blocks as shown in Figure A–2 or Figure A–3.
4. Re-connect the probe.
5. Re-connect the ribbon cable to the keypad.
6. Close the analyzer enclosure door.
Installing/Replacing the Pressure Sensor
The pressure sensor is optional on the OXY5500 analyzer. Use this procedure
to install or replace the pressure sensor.
Refer to the procedure called “Installing the pressure sensor” on page B-9
and to Table A–2 on page A–6 for the pressure sensor kit part number to install
this option.
Figure B–5 Removing the ground cable