OXY5500 Optical Oxygen Analyzer BA02195C
4
–16 Endress+Hauser
• Measurement Name: This box displays the selected measurement
file in which all data is stored when logging is turned on.
• General: Here the Sensor type of the currently connected oxygen
sensor is displayed.
– The currently measured or manually set Pressure value is
also displayed in the General box. For automatic measurement,
the display will show the interpreted pressure value from the
4-20mA input
.
– In the lower right of the General box, the time Interval at which
the measurements are taken is displayed.
– Next provides the time period (the countdown during a running
measurement) until the next measurement.
– The RATA is displayed at the bottom of the screen.
– Error Codes are also displayed under the General box. Refer to
Table 4–1. Error codes are also logged together with
measurement data. During error-free measurements, a 0 value
will display.
• Press to return to Simple view.
• Press to display the measurement graph. Pressing the button
will switch to the graphical presentation of the current
measurements. Refer to “Graph screen” on page 4-17.
• Press to return to the MAIN MENU screen.
Error codes
The error code is a bit combination of multiple errors. Table 4–1 shows a list of
error bits. For more information on error notifications, refer to Table B–3 on
page B–18. Some examples of error codes are shown below:
“Error code: 1” = No RTD (Pt100) (bit O)
The temperature unit can be changed in Manual mode. Values
ranging from -99
° C to 199° C can be input into the MEAS.
SETTINGS window. Refer to “Temperature compensation” on
page 4-19.
The measurement file can be changed in the MEAS. SETTINGS
menu. Refer to “Logging and data management” on page
4-23.
If no pressure sensor is connected, the display will read 1013 mbar.