EasyManua.ls Logo

Endress+Hauser OXY5500 - To Enter Editing Mode; To Exit Editing Mode; Changing the Sensor Type

Endress+Hauser OXY5500
146 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
BA02195C OXY5500 Optical Oxygen Analyzer
Endress+Hauser 4
31
Select Yes to stop the measurement in order to display the SENSOR TYPE AND
SENSOR CONSTANTS window. Refer to Figure 4–33.
Use the arrow buttons to navigate between input fields.
To enter editing mode
1. Press to edit the highlighted field.
2. Change the setting or value (one digit at a time) by pressing the
buttons.
3. Make the desired change to an input field.
4. Press again to save the changes.
To exit editing mode
1. Press to cancel and exit.
Changing the sensor type
If it is necessary to change the probe type in the field, change the sensor type
(OP-3, OP-6 or OP-9) according to the sensor that is connected to the analyzer.
The displayed sensor constants (dKSV1, dKSV2, dPhi1, dPhi2, f1 and m) will
change according to the sensor type selected.
The sensor constant values can also be located on the Calibration
Certificate delivered with the optical oxygen sensor. Refer to the
example in Figure 4–34.
Figure 433 Sensor type and sensor constants
- Sensor type menu selected for editing

Table of Contents

Other manuals for Endress+Hauser OXY5500

Related product manuals