Diagnostics and troubleshooting Proline Prowirl F 200 PROFIBUS PA
146 Endress+Hauser
Diagnostic information Remedy instructions Influenced measured
variables
No. Short text
046 Sensor limit exceeded 1. Check plug connections
2. Change pre-amplifier
3. Change DSC sensor
• Calculated saturated
steam pressure
• Density
• Energy flow
• Flow velocity
• Heat flow difference
• Low flow cut off
• Mass flow
• Total mass flow
• Switch output status
• Pressure
• Reynolds number
• Specific volume
• Corrected volume flow
• Steam quality
• Degrees of superheat
• Volume flow
Measured variable status
Quality Good
Quality substatus Maintenance demanded
Coding (hex) 0xA8 to 0xAB
Status signal S
Diagnostic behavior Warning
Diagnostic information Remedy instructions Influenced measured
variables
No. Short text
062 Sensor connection defective 1. Check plug connections
2. Change pre-amplifier
3. Change DSC sensor
• Calculated saturated
steam pressure
• Density
• Energy flow
• Flow velocity
• Heat flow difference
• Low flow cut off
• Mass flow
• Total mass flow
• Switch output status
• Pressure
• Reynolds number
• Specific volume
• Corrected volume flow
• Steam quality
• Degrees of superheat
• Temperature
• Volume flow
Measured variable status
Quality Bad
Quality substatus Maintenance alarm
Coding (hex) 0x24 to 0x27
Status signal F
Diagnostic behavior Alarm
Diagnostic information Remedy instructions Influenced measured
variables
No. Short text
082 Data storage 1. Change main electronic module
2. Change sensor
• Calculated saturated
steam pressure
• Density
• Energy flow
• Flow velocity
• Heat flow difference
• Low flow cut off
• Mass flow
• Total mass flow
• Switch output status
• Pressure
• Reynolds number
• Specific volume
• Corrected volume flow
• Steam quality
• Degrees of superheat
• Temperature
• Volume flow
Measured variable status
Quality Bad
Quality substatus Maintenance alarm
Coding (hex) 0x24 to 0x27
Status signal F
Diagnostic behavior Alarm