Commissioning Proline Promag H 300 PROFINET
124 Endress+Hauser
Parameter Prerequisite Description Selection / User
entry / User
interface
Factory setting
Value current input 1 to n In the Current input 1 to n
simulation parameter, the On
option is selected.
Enter the current value for
simulation.
0 to 22.5 mA 0 mA
Current output 1 to n simulation Switch the simulation of the
current output on and off.
Off
On
Off
Value current output 1 to n In the Current output 1 to n
simulation parameter, the On
option is selected.
Enter the current value for
simulation.
3.59 to 22.5 mA 3.59 mA
Frequency output simulation 1 to n In the Operating mode
parameter, the Frequency
option is selected.
Switch the simulation of the
frequency output on and off.
Off
On
Off
Frequency value 1 to n In the Frequency output
simulation 1 to n parameter,
the On option is selected.
Enter the frequency value for
the simulation.
0.0 to 12 500.0 Hz 0.0 Hz
Pulse output simulation 1 to n In the Operating mode
parameter, the Pulse option is
selected.
Set and switch off the pulse
output simulation.
For Fixed value option:
Pulse width parameter
(→  96) defines the
pulse width of the pulses
output.
Off
Fixed value
Down-counting
value
Off
Pulse value 1 to n In the Pulse output
simulation 1 to n parameter,
the Down-counting value
option is selected.
Enter the number of pulses for
simulation.
0 to 65 535 0
Switch output simulation 1 to n In the Operating mode
parameter, the Switch option
is selected.
Switch the simulation of the
switch output on and off.
Off
On
Off
Switch status 1 to n Select the status of the status
output for the simulation.
Open
Closed
Open
Relay output 1 to n simulation Switch simulation of the relay
output on and off.
Off
On
Off
Switch status 1 to n The On option is selected in
the Switch output simulation
1 to n parameter parameter.
Select status of the relay
output for the simulation.
Open
Closed
Open
Pulse output simulation Set and switch off the pulse
output simulation.
For Fixed value option:
Pulse width parameter
defines the pulse width
of the pulses output.
Off
Fixed value
Down-counting
value
Off
Pulse value In the Pulse output
simulation parameter, the
Down-counting value option
is selected.
Set and switch off the pulse
output simulation.
0 to 65 535 0
Device alarm simulation Switch the device alarm on and
off.
Off
On
Off
Diagnostic event category Select a diagnostic event
category.
Sensor
Electronics
Configuration
Process
Process