Proline Promag H 300 PROFINET Commissioning
Endress+Hauser 117
Navigation
"Setup" menu → Advanced setup → Electrode cleaning circuit
‣
Electrode cleaning circuit
Electrode cleaning circuit
→ 117
ECC duration
→ 117
ECC recovery time
→ 117
ECC cleaning cycle
→ 117
ECC Polarity
→ 117
Parameter overview with brief description
Parameter Prerequisite Description Selection / User
entry / User
interface
Factory setting
Electrode cleaning circuit For the following order code:
"Application package", option
EC "ECC electrode cleaning"
Enable the cyclic electrode
cleaning circuit.
• Off
• On
Off
ECC duration For the following order code:
"Application package", option
EC "ECC electrode cleaning"
Enter the duration of electrode
cleaning in seconds.
0.01 to 30 s 2 s
ECC recovery time For the following order code:
"Application package", option
EC "ECC electrode cleaning"
Define recovery time after
electrode cleaning. During this
time the current output values
will be held at last valid value.
1 to 600 s 60 s
ECC cleaning cycle For the following order code:
"Application package", option
EC "ECC electrode cleaning"
Enter the pause duration
between electrode cleaning
cycles.
0.5 to 168 h 0.5 h
ECC Polarity For the following order code:
"Application package", option
EC "ECC electrode cleaning"
Select the polarity of the
electrode cleaning circuit.
• Positive
• Negative
Depends on the
electrode material:
• Platinum:
Negative option
• Tantalum, Alloy
C22, stainless
steel: Positive
option