Diagnostics and troubleshooting Proline Promag H 300 PROFINET
154 Endress+Hauser
12.7.2 Diagnostic of electronic
Diagnostic information Remedy instructions Influenced measured
variables
No. Short text
201 Device failure 1. Restart device
2. Contact service
Conductivity
Corrected conductivity
Measured values 1
Measured values 2
Measured values 3
Density
Electronic temperature
Flow velocity
Mass flow
Coil current shot time
Reference electrode
potential against PE
Noise
Reference density
Corrected volume flow
Temperature
Status
Volume flow
Measured variable status
Quality Bad
Quality substatus Maintenance alarm
Coding (hex) 0x24 to 0x27
Status signal F
Diagnostic behavior Alarm
Diagnostic information Remedy instructions Influenced measured
variables
No. Short text
242 Software incompatible 1. Check software
2. Flash or change main electronics
module
Conductivity
Corrected conductivity
Measured values 1
Measured values 2
Measured values 3
Density
Electronic temperature
Flow velocity
Mass flow
Coil current shot time
Reference electrode
potential against PE
Noise
Reference density
Corrected volume flow
Temperature
Status
Volume flow
Measured variable status
Quality Good
Quality substatus Ok
Coding (hex) 0x80 to 0x83
Status signal F
Diagnostic behavior Alarm
Diagnostic information Remedy instructions Influenced measured
variables
No. Short text
252 Modules incompatible 1. Check electronic modules
2. Check if correct modules are available
(e.g. NEx, Ex)
3. Replace electronic modules
Conductivity
Corrected conductivity
Measured values 1
Measured values 2
Measured values 3
Density
Electronic temperature
Flow velocity
Mass flow
Coil current shot time
Reference electrode
potential against PE
Noise
Reference density
Corrected volume flow
Temperature
Volume flow
Measured variable status
Quality Good
Quality substatus Ok
Coding (hex) 0x80 to 0x83
Status signal F
Diagnostic behavior Alarm