Technical data Proline Promag H 300 PROFINET
198 Endress+Hauser
Temperature
±0.5 °C (±0.9 °F)
Electrical conductivity
• Max. ±5 % o.r.
• Max. ±1 % o.r. for DN 15 to 150 in conjunction with stainless steel process connections,
1.4404 (F316L)
Temperature measurement
response time
T
90
< 15 s
Influence of ambient
temperature
Current output
Temperature coefficient Max. 1 μA/°C
Pulse/frequency output
Temperature coefficient No additional effect. Included in accuracy.
16.7 Installation
Chapter "Mounting requirements" → 21
16.8 Environment
Ambient temperature
range
→ 23
Temperature tables
Observe the interdependencies between the permitted ambient and fluid
temperatures when operating the device in hazardous areas.
For detailed information on the temperature tables, see the separate document
entitled "Safety Instructions" (XA) for the device.
Storage temperature The storage temperature corresponds to the operating temperature range of the
transmitter and the sensor → 23.
• Protect the measuring device against direct sunlight during storage in order to avoid
unacceptably high surface temperatures.
• Select a storage location where moisture cannot collect in the measuring device as
fungus or bacteria infestation can damage the liner.
• If protection caps or protective covers are mounted these should never be removed
before installing the measuring device.
Degree of protection Measuring device
• As standard: IP66/67, type 4X enclosure
• When housing is open: IP20, type 1 enclosure
• Display module: IP20, type 1 enclosure
• With the order code for "Sensor options", option CM: IP69 can also be ordered
External WLAN antenna
IP67