Proline Promag H 300 PROFINET Technical data
Endress+Hauser 187
16 Technical data
16.1 Application
The measuring device is only suitable for flow measurement of liquids with a minimum
conductivity of 5 µS/cm.
Depending on the version ordered, the measuring device can also measure potentially
explosive, flammable, poisonous and oxidizing media.
To ensure that the device remains in proper operating condition for its service life, use the
measuring device only for media against which the process-wetted materials are
sufficiently resistant.
16.2 Function and system design
Measuring principle Electromagnetic flow measurement on the basis of Faraday's law of magnetic induction.
Measuring system The device consists of a transmitter and a sensor.
The device is available as a compact version:
The transmitter and sensor form a mechanical unit.
For information on the structure of the device → 14
16.3 Input
Measured variable Direct measured variables
• Volume flow (proportional to induced voltage)
• Temperature
1)
• Electrical conductivity
Calculated measured variables
• Mass flow
• Corrected volume flow
• Corrected electrical conductivity
1)
Measuring range Typically v = 0.01 to 10 m/s (0.03 to 33 ft/s) with the specified accuracy
Flow characteristic values in SI units: DN 2 to 125 (¹⁄₁₂ to 5")
Nominal diameter
Recommended
flow
Factory settings
min./max. full
scale value
(v ~ 0.3/10 m/s)
Full scale value
current output
(v ~ 2.5 m/s)
Pulse value
(~ 2 pulse/s)
Low flow cut off
(v ~ 0.04 m/s)
[mm] [in] [dm
3
/min] [dm
3
/min] [dm
3
] [dm
3
/min]
2 ¹⁄₁₂ 0.06 to 1.8 0.5 0.005 0.01
4 ⁵⁄₃₂ 0.25 to 7 2 0.025 0.05
1) Available only for nominal diameters DN 15 to 150 (½ to 6") and with the order code for "Sensor option", option CI "Medium temperature
measurement".