EasyManua.ls Logo

Endress+Hauser Proline Promag H 300 - Configuration Management

Endress+Hauser Proline Promag H 300
220 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Proline Promag H 300 PROFINET Commissioning
Endress+Hauser 119
Parameter Prerequisite Description Selection / User
entry / User
interface
Factory setting
WLAN password Enter WLAN password.
WLAN IP address Enter IP address of the WLAN
interface of the device.
4 octet: 0 to 255 (in
the particular octet)
192.168.1.212
WLAN passphrase The WPA2-PSK option is
selected in the Security type
parameter.
Enter the network key (8 to 32
characters).
The network key
supplied with the device
should be changed
during commissioning
for security reasons.
8 to 32-digit
character string
comprising numbers,
letters and special
characters (without
spaces)
Serial number of the
measuring device
(e.g. L100A802000)
Assign SSID name Select which name will be used
for SSID: device tag or user-
defined name.
Device tag
User-defined
User-defined
SSID name The User-defined option is
selected in the Assign SSID
name parameter.
The WLAN access point
option is selected in the
WLAN mode parameter.
Enter the user-defined SSID
name (max. 32 characters).
The user-defined SSID
name may only be
assigned once. If the SSID
name is assigned more
than once, the devices
can interfere with one
another.
Max. 32-digit
character string
comprising numbers,
letters and special
characters
EH_device
designation_last 7
digits of the serial
number (e.g.
EH_Promag_300_A
802000)
Connection state Displays the connection status. Connected
Not connected
Not connected
Received signal strength Shows the received signal
strength.
Low
Medium
High
High
* Visibility depends on order options or device settings
10.6.7 Configuration management
After commissioning, you can save the current device configurationor restore the previous
device configuration.
You can do so using the Configuration management parameter and the related options
found in the Configuration backup submenu.
Navigation
"Setup" menu → Advanced setup → Configuration backup
Configuration backup
Operating time
→  120
Last backup
→  120
Configuration management
→  120
Backup state
→  120
Comparison result
→  120

Table of Contents

Other manuals for Endress+Hauser Proline Promag H 300

Related product manuals