Proline Promag H 300 PROFINET Diagnostics and troubleshooting
Endress+Hauser 171
Diagnostic information Remedy instructions Influenced measured
variables
No. Short text
594 Relay output simulation Deactivate simulation switch output
Measured variable status
Quality Good
Quality substatus Ok
Coding (hex) 0x80 to 0x83
Status signal C
Diagnostic behavior Warning
12.7.4 Diagnostic of process
Diagnostic information Remedy instructions Influenced measured
variables
No. Short text
803 Current loop 1. Check wiring
2. Change I/O module
Measured variable status
Quality Good
Quality substatus Ok
Coding (hex) 0x80 to 0x83
Status signal F
Diagnostic behavior Alarm
Diagnostic information Remedy instructions Influenced measured
variables
No. Short text
832 Electronic temperature too high Reduce ambient temperature Conductivity
Corrected conductivity
Measured values 1
Measured values 2
Measured values 3
Density
Electronic temperature
Flow velocity
Mass flow
Coil current shot time
Reference electrode
potential against PE
Noise
Reference density
Corrected volume flow
Temperature
Status
Volume flow
Measured variable status [from the factory]
1)
Quality Good
Quality substatus Ok
Coding (hex) 0x80 to 0x83
Status signal S
Diagnostic behavior Warning
1) Diagnostic behavior can be changed. This causes the overall status of the measured variable to change.