Proline Promag H 300 PROFINET Diagnostics and troubleshooting
Endress+Hauser 171
Diagnostic information Remedy instructions Influenced measured
variables
No. Short text
594 Relay output simulation Deactivate simulation switch output –
Measured variable status
Quality Good
Quality substatus Ok
Coding (hex) 0x80 to 0x83
Status signal C
Diagnostic behavior Warning
12.7.4 Diagnostic of process
Diagnostic information Remedy instructions Influenced measured
variables
No. Short text
803 Current loop 1. Check wiring
2. Change I/O module
–
Measured variable status
Quality Good
Quality substatus Ok
Coding (hex) 0x80 to 0x83
Status signal F
Diagnostic behavior Alarm
Diagnostic information Remedy instructions Influenced measured
variables
No. Short text
832 Electronic temperature too high Reduce ambient temperature • Conductivity
• Corrected conductivity
• Measured values 1
• Measured values 2
• Measured values 3
• Density
• Electronic temperature
• Flow velocity
• Mass flow
• Coil current shot time
• Reference electrode
potential against PE
• Noise
• Reference density
• Corrected volume flow
• Temperature
• Status
• Volume flow
Measured variable status [from the factory]
1)
Quality Good
Quality substatus Ok
Coding (hex) 0x80 to 0x83
Status signal S
Diagnostic behavior Warning
1) Diagnostic behavior can be changed. This causes the overall status of the measured variable to change.