Diagnostics and troubleshooting Proline Promag H 300 PROFINET
170 Endress+Hauser
Diagnostic information Remedy instructions Influenced measured
variables
No. Short text
520 I/O 1 to n hardware configuration invalid 1. Check I/O hardware configuration
2. Replace wrong I/O module
3. Plug the module of double pulse output
on correct slot
Measured variable status
Quality Good
Quality substatus Ok
Coding (hex) 0x80 to 0x83
Status signal F
Diagnostic behavior Alarm
Diagnostic information Remedy instructions Influenced measured
variables
No. Short text
530 Electrode cleaning is running Turn off ECC Conductivity
Corrected conductivity
Density
Electronic temperature
Flow velocity
Mass flow
Coil current shot time
Reference electrode
potential against PE
Noise
Reference density
Corrected volume flow
Temperature
Status
Volume flow
Measured variable status
Quality Good
Quality substatus Function check
Coding (hex) 0xBC to 0xBF
Status signal C
Diagnostic behavior Warning
Diagnostic information Remedy instructions Influenced measured
variables
No. Short text
531 Empty pipe adjustment faulty Execute EPD adjustment Conductivity
Corrected conductivity
Flow velocity
Mass flow
Coil current shot time
Reference electrode
potential against PE
Noise
Corrected volume flow
Status
Volume flow
Measured variable status [from the factory]
1)
Quality Good
Quality substatus Ok
Coding (hex) 0x80 to 0x83
Status signal S
Diagnostic behavior Warning
1) Diagnostic behavior can be changed. This causes the overall status of the measured variable to change.
Diagnostic information Remedy instructions Influenced measured
variables
No. Short text
537 Configuration 1. Check IP addresses in network
2. Change IP address
Measured variable status
Quality Good
Quality substatus Ok
Coding (hex) 0x80 to 0x83
Status signal F
Diagnostic behavior Warning