Proline Promag H 300 PROFINET Diagnostics and troubleshooting
Endress+Hauser 169
Diagnostic information Remedy instructions Influenced measured
variables
No. Short text
494 Switch output simulation 1 to n Deactivate simulation switch output –
Measured variable status
Quality Good
Quality substatus Ok
Coding (hex) 0x80 to 0x83
Status signal C
Diagnostic behavior Warning
Diagnostic information Remedy instructions Influenced measured
variables
No. Short text
495 Diagnostic event simulation Deactivate simulation –
Measured variable status
Quality Good
Quality substatus Ok
Coding (hex) 0x80 to 0x83
Status signal C
Diagnostic behavior Warning
Diagnostic information Remedy instructions Influenced measured
variables
No. Short text
496 Status input simulation Deactivate simulation status input –
Measured variable status
Quality Good
Quality substatus Ok
Coding (hex) 0x80 to 0x83
Status signal C
Diagnostic behavior Warning
Diagnostic information Remedy instructions Influenced measured
variables
No. Short text
511 ISEM settings faulty 1. Check measuring period and integration
time
2. Check sensor properties
• Conductivity
• Corrected conductivity
• Density
• Electronic temperature
• Flow velocity
• Mass flow
• Coil current shot time
• Reference electrode
potential against PE
• Noise
• Reference density
• Corrected volume flow
• Temperature
• Status
• Volume flow
Measured variable status
Quality Bad
Quality substatus Maintenance alarm
Coding (hex) 0x24 to 0x27
Status signal C
Diagnostic behavior Alarm