Proline Promag H 300 PROFINET Diagnostics and troubleshooting
Endress+Hauser 155
Diagnostic information Remedy instructions Influenced measured
variables
No. Short text
252 Modules incompatible 1. Check if correct electronic modul is
plugged
2. Replace electronic module
Conductivity
Corrected conductivity
Measured values 1
Measured values 2
Measured values 3
Density
Electronic temperature
Flow velocity
Mass flow
Coil current shot time
Reference electrode
potential against PE
Noise
Reference density
Corrected volume flow
Temperature
Status
Volume flow
Measured variable status
Quality Bad
Quality substatus Maintenance alarm
Coding (hex) 0x24 to 0x27
Status signal F
Diagnostic behavior Alarm
Diagnostic information Remedy instructions Influenced measured
variables
No. Short text
262 Sensor electronic connection faulty 1. Check or replace connection cable
between sensor electronic module
(ISEM) and main electronics
2. Check or replace ISEM or main
electronics
Conductivity
Corrected conductivity
Measured values 1
Measured values 2
Measured values 3
Density
Electronic temperature
Flow velocity
Mass flow
Coil current shot time
Reference electrode
potential against PE
Noise
Reference density
Corrected volume flow
Temperature
Status
Volume flow
Measured variable status
Quality Bad
Quality substatus Maintenance alarm
Coding (hex) 0x24 to 0x27
Status signal F
Diagnostic behavior Alarm
Diagnostic information Remedy instructions Influenced measured
variables
No. Short text
270 Main electronic failure Change main electronic module Conductivity
Corrected conductivity
Measured values 1
Measured values 2
Measured values 3
Density
Electronic temperature
Flow velocity
Mass flow
Coil current shot time
Reference electrode
potential against PE
Noise
Reference density
Corrected volume flow
Temperature
Status
Volume flow
Measured variable status
Quality Bad
Quality substatus Maintenance alarm
Coding (hex) 0x24 to 0x27
Status signal F
Diagnostic behavior Alarm