Diagnostics and troubleshooting Proline Promag H 300 PROFINET
156 Endress+Hauser
Diagnostic information Remedy instructions Influenced measured
variables
No. Short text
271 Main electronic failure 1. Restart device
2. Change main electronic module
Conductivity
Corrected conductivity
Measured values 1
Measured values 2
Measured values 3
Density
Electronic temperature
Flow velocity
Mass flow
Coil current shot time
Reference electrode
potential against PE
Noise
Reference density
Corrected volume flow
Temperature
Status
Volume flow
Measured variable status
Quality Bad
Quality substatus Maintenance alarm
Coding (hex) 0x24 to 0x27
Status signal F
Diagnostic behavior Alarm
Diagnostic information Remedy instructions Influenced measured
variables
No. Short text
272 Main electronic failure 1. Restart device
2. Contact service
Conductivity
Corrected conductivity
Measured values 1
Measured values 2
Measured values 3
Density
Electronic temperature
Flow velocity
Mass flow
Coil current shot time
Reference electrode
potential against PE
Noise
Reference density
Corrected volume flow
Temperature
Status
Volume flow
Measured variable status
Quality Bad
Quality substatus Maintenance alarm
Coding (hex) 0x24 to 0x27
Status signal F
Diagnostic behavior Alarm
Diagnostic information Remedy instructions Influenced measured
variables
No. Short text
273 Main electronic failure Change electronic Conductivity
Corrected conductivity
Measured values 1
Measured values 2
Measured values 3
Density
Electronic temperature
Flow velocity
Mass flow
Coil current shot time
Reference electrode
potential against PE
Noise
Reference density
Corrected volume flow
Temperature
Status
Volume flow
Measured variable status
Quality Bad
Quality substatus Maintenance alarm
Coding (hex) 0x24 to 0x27
Status signal F
Diagnostic behavior Alarm