Diagnostics and troubleshooting Proline Promag H 300 PROFINET
152 Endress+Hauser
Diagnostic information Remedy instructions Influenced measured
variables
No. Short text
083 Memory content 1. Restart device
2. Restore HistoROM S-DAT backup
('Device reset' parameter)
3. Replace HistoROM S-DAT
Conductivity
Corrected conductivity
Measured values 1
Measured values 2
Measured values 3
Density
Electronic temperature
Flow velocity
Mass flow
Coil current shot time
Reference electrode
potential against PE
Noise
Reference density
Corrected volume flow
Temperature
Status
Volume flow
Measured variable status
Quality Bad
Quality substatus Maintenance alarm
Coding (hex) 0x24 to 0x27
Status signal F
Diagnostic behavior Alarm
Diagnostic information Remedy instructions Influenced measured
variables
No. Short text
168 Coating detected Clean measuring tube
Measured variable status
Quality Good
Quality substatus Ok
Coding (hex) 0x80 to 0x83
Status signal M
Diagnostic behavior Warning
Diagnostic information Remedy instructions Influenced measured
variables
No. Short text
169 Conductivity measurement failed 1. Check grounding conditions
2. Deactivate conductivity measurement
Conductivity
Corrected conductivity
Electronic temperature
Flow velocity
Mass flow
Coil current shot time
Reference electrode
potential against PE
Noise
Corrected volume flow
Temperature
Status
Volume flow
Measured variable status
Quality Good
Quality substatus Ok
Coding (hex) 0x80 to 0x83
Status signal M
Diagnostic behavior Warning