EasyManua.ls Logo

Endress+Hauser Proline Promag H 300 - Page 203

Endress+Hauser Proline Promag H 300
220 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Proline Promag H 300 PROFINET Technical data
Endress+Hauser 203
External WLAN antenna
Antenna: ASA plastic (acrylic ester-styrene-acrylonitrile) and nickel-plated brass
Adapter: Stainless steel and nickel-plated brass
Cable: Polyethylene
Plug: Nickel-plated brass
Angle bracket: Stainless steel
Grounding rings
Standard: 1.4435 (316L)
Optional: Alloy C22, tantalum
Wall mounting kit
Stainless steel 1.4301 (304)
Spacer
1.4435 (F316L)
Fitted electrodes 2 measuring electrodes for signal detection
1 empty pipe detection electrode for empty pipe detection/temperature measurement
(only DN 15 to 150 (½ to 6"))
Process connections With O-ring seal:
Welding nipple (DIN EN ISO 1127, ODT/SMS, ISO 2037)
Flange (EN (DIN), ASME, JIS)
Flange from PVDF (EN (DIN), ASME, JIS)
External thread
Internal thread
Hose connection
PVC adhesive sleeve
With aseptic molded seal:
Coupling (DIN 11851, DIN 11864-1, ISO 2853, SMS 1145)
Flange DIN 11864-2
For information on the different materials used in the process connections →  202
Surface roughness Stainless steel electrodes, 1.4435 (316L); Alloy C22, 2.4602 (UNS N06022); platinum;
tantalum:
≤ 0.3 to 0.5 µm (11.8 to 19.7 µin)
(All data relate to parts in contact with fluid)
Liner with PFA:
≤ 0.4 µm (15.7 µin)
(All data relate to parts in contact with fluid)
Stainless steel process connections:
With O-ring seal: ≤ 1.6 µm (63 µin)
With aseptic seal: ≤ 0.8 µm (31.5 µin)
Optional: ≤ 0.38 µm (15 µin)
(All data relate to parts in contact with fluid)