Electrical connection Proline Promag H 300 PROFINET
44 Endress+Hauser
Setting the device name using the DIP switches
The last part of the device name can be set using DIP switches 1-8. The address range is
between 1 and 254 (factory setting: serial number of the device )
Overview of the DIP switches
DIP switch Bit Description
1 128
Configurable part of the device name
2 64
3 32
4 16
5 8
6 4
7 2
8 1
Example: Setting the device name EH-PROMAG300-065
DIP switch ON/OFF Bit Device name
1 OFF –
2 ON 64
3…7 OFF –
8 ON 1
Serial number of the device: 065
EH-PROMAG300-065
Setting the device name
Risk of electric shock when opening the transmitter housing.
‣
Before opening the transmitter housing:
‣
Disconnect the device from the power supply.
The default IP address may not be activated → 45.
1
2
3
4
5
6
7
8
128
64
32
16
8
4
2
1
Name of Station
A0034498
1. Depending on the housing version, loosen the securing clamp or fixing screw of the
housing cover.
2. Depending on the housing version, unscrew or open the housing cover and
disconnect the local display from the main electronics module where necessary .
3. Set the desired device name using the corresponding DIP switches on the
I/O electronics module.
4. Reverse the removal procedure to reassemble the transmitter.