ManagEMEnt > protoCoLs
biCyCLE
Protocol Type ‘SpO2-controlled‘
Subsequently, the SpO2 value is checked
on the basis of the Workload Control pa-
rameters. If another deviation is detect-
ed, the training load will automatically
be adjusted.
Entries can be made as absolute or rel-
ative values with respect to the training
load.
If you do not enter any values in the SpO2
Control window, the system will use the
preset defaults.
Protocol Type ‘Interval HR-controlled‘
In the Training pane click [ Extended ] to
open the Interval HR Control window
with the Workload Control parameters.
In this window, enter the permitted de-
viation at Threshold [1/min] and the time
interval at Delay [s]. If the heart rate
deviates for the entire time interval, the
workload will be adjusted accordingly.
At Type, you determine the nature of the
workload change: absolute or relative.