5500A
Operator Manual
6-18
Features
Temperature Standard
its-90
Display Contrast*
level 7,7
Host Connection
gpib (IEEE-488)
Display Brightness*
level 1,0
GPIB Port Address
4
RTD Power Up
Default Type
pt385
Serial Ports
8 bits, 1 stop bit, xon/xoff, parity
none, 9600 baud, wait 30 sec.
Thermocouple Power
Up Default Type
K
EOL (end of line)
CRLF
Source Preference
5500
EOF (end of file)
012,000
Current Limits
+11 A
Remote I/F
term
Voltage Limits
+1000 V
Remote commands (see Chapter 6)
SRQSTR
SRQ: %02x %02x %04x %04x
*PUD string
cleared
* Output Display and Control Display, respectively. There are 8 levels: 0,1,2,3,4,5,6,7.
FUNC?
Sequential Overlapped
Coupled
IEEE-488 RS-232
x x xx
x
(Function query) Return the present output, measurement, or calibration function. See
the response below for output & measurement modes.
Responses: DCV (dc volts function)
ACV (ac volts function)
DCI (dc current function)
ACI (ac current function)
RES (ohms function)
CAP (capacitance function)
RTD (temperature with an rtd function)
TC_OUT (temperature with a thermocouple function)
DC_POWER (dc power function)
AC_POWER (ac power function)
DCV_DCV (dual dc volts function)
ACV_ACV (dual ac volts function)
TC_MEAS (measure temperature with a thermocouple)
SACV (oscilloscope ac volts function)
SDCV (oscilloscope dc volts function)
MARKER (oscilloscope marker function)
LEVSINE (oscilloscope leveled sine function)
EDGE (oscilloscope edge function)
Example: FUNC? returns DCV_DCV
Return DCV_DCV when the 5500A Calibrator output function dual dc volts.