Remote Operation
Commands 5
5-15
Table 5-1. Command Summary by Function (continued)
RS-232-C Interface Parameter Setting Commands
SP_EOF Sets the End-Of-File (EOF) string
SP_EOF? Returns the End-Of-File (EOF) string
SP_SET Sets baud, terminal or computer mode, stall method, data bits, stop bits, parity,
and End-Of-Line (EOL) string
SP_SET? Returns baud rate, terminal or computer mode, stall method, data bits, stop
bits, parity, and End-Of-Line (EOL) string
SP_TIME Sets the timeout period
SP_TIME? Returns the timeout period
Calibration, Testing, and Diagnostics Commands for the Calibrator
CAL_ADJ Does the internal portion of calibration
CAL_CHK Starts calibration check
CAL_CLST? Returns a group of calibration constant names and their values
CAL_CONF Sets the calibration specifications to a confidence level of either 95% or 99%.
CAL_CONF? Returns the current calibration confidence level.
CAL_CONST? Returns the value of a particular constant
CAL_DATE? Returns the date of the most recent calibration
CAL_DAYS? Returns the days elapsed since last calibration recall
CAL_INTV Sets the calibration interval
CAL_INTV? Returns the calibration interval
CAL_PR Prints a calibration report out the serial interface
CAL_REF Does a calibration procedure using a 1, 10 k, or 10V standard
CAL_RNG Starts a self calibration procedure of a range
CAL_RPT? Returns a formatted calibration report
CAL_SHIFT? Returns the shift of a range due to a calibration action
CAL_SLST? Returns the shifts of a group of ranges
CAL_STORE Stores new calibration constants in nonvolatile memory
CAL_TEMP Sets the temperature for calibration
CAL_TEMP? Returns the calibration temperature last specified
CAL_WBFLAT Does Wideband AC Module (Option 5700A-03) flatness calibration
CAL_WBGAIN Does Wideband AC Module (Option 5700A-03) gain calibration
CAL_ZERO Does internal zeros calibration
DIAG Runs self diagnostics