-39-
4.9.2.2. Setting the preset value
Description
Preset value: Value which appears on the total counter when the total value has been reset.
Settable range:0 to 99999999
<Note> A resetting action simultaneously resets both forward total memory and reverse total memory.
Set the total unit beforehand in the MEASURE SETUP mode. (Refer to 4.8.3.)
When setting, stop status is set at the total mode.
Total value
Preset value
Time
Reset
Reset
Reset
For actual keying, refer to the typical operation indicated below. Set the protection to OFF beforehand. (See Section 4.4.1.)
Operation
(example)
Set the preset value to 100m
3
.
* Set the total unit beforehand.
Key operation Description Display
▼
Press the key twice to display “OUTPUT SETUP”.
OUTPUT SETUP
ENT
▼
Press the
ENT
key once to display “ZERO ADJUSTMENT”.
ZERO ADJUSTMENT
SET ZERO
▼
Press the key for 5 times to display “TOTAL”.
TOTAL
ENT
▼
Press the
ENT
key once to display “TOTAL MODE”.
TOTAL MODE
STOP
▼
Press the key twice to display “TOTAL PRESET”
TOTAL PRESET
0 m3
ENT
▼
Press the
ENT
key once to display the cursor.
TOTAL PRESET
0000000
▼
Press the key for 6 times to move the cursor.
* Note that, it cannot be entered on the first digit (leftmost).
TOTAL PRESET
00000000
m3
▼
Press the key once to display “1”.
TOTAL PRESET
00000100 m3
ENT
▼
▼
▼
▼
Press the
ENT
key once to register.
TOTAL PRESET
COMPLETE
↓
――― “ TOTAL PRESET” has been registered. ―――
TOTAL PRESET
100 m3
▼
Press the key for 4 times to display “TOTAL MODE”.
TOTAL MODE
STOP
ENT
▼
Press the
ENT
key once to blink the cursor.
TOTAL MODE
STOP
▼
Press the key once, and select “TOTAL PRESET”.
TOTAL MODE
TOTAL RESET
ENT
▼
▼
▼
▼
Press the
ENT
key once to register.
TOTAL MODE
COMPLETE
↓
――― “TOTAL MODE” has been registered. ―――
TOTAL MODE
TOTAL RESET
ESC
Press the
ESC
key twice and then press the key for 3 times to enter
the measurement mode.
0.000 m/s
0.000 m3/h