EasyManua.ls Logo

Fuji Electric Time Delta-C FSV-2 - Installation Location Cautions

Fuji Electric Time Delta-C FSV-2
119 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
-v-
CAUTION ON INSTALLATION LOCATION
(1) A place that provides enough space for periodic inspection and wiring work.
(2) A place not exposed to direct sunshine nor inclement weather.
(3) A place free from excessive vibration, dust, dirt and moisture.
(4) A place not subjected to radiated heat from a heating furnace, etc.
(5) A place not subjected to corrosive atmosphere.
(6) A place not to be submerged.
(7) A place remote from electrical devices (motor, transformer, etc.) which generate electromagnetic induction
noise, electrostatic noise, etc.
(8) A place not subjected to excessive fluid pulsation such as pump discharge side.
(9) A place that provides enough place for the length of the straight pipe.
(10) A place where ambient temperature and humidity are -20 to +55°C and 95% RH or less for flow transmitter
(FSV).
(11) Altitude: up to 2000m
(12) Installation category: II
(13) Pollution degree: 2
CAUTION

Table of Contents

Related product manuals