4.4. Parameter protection
4.4.1. Protection ON/OFF
Description
Parameters can be protected so that the flow meter settings will not carelessly be changed.
Parameters can be protected by setting the "ID No." (Note) in the maintenance mode.
Note) 4 digits are factory set at "0000". (Refer to Section 4.11.7.)
Setting range: PROTECTION ON : Parameter cannot be changed.
PROTECTION OFF : Parameter can be changed.
* 1 hour after “PROTECTION OFF” is set, “PROTECTION ON” is automatically set.
* Protection is set after turning power on.
For actual keying, refer to the typical operation indicated below.
Operation
(example)
Change the protection from ON to OFF (suppose ID No. is "2234").
Key operation Description Display
Press the key in the measurement mode once to indicate “PAR.
PROTECTION”.
PAR.PROTECT
PROTECTION ON
ENT
Press the
ENT
key once to blink the 2nd line.
PAR.PROTECT
PROTECTION ON
Press the key once to display "PROTECTION OFF".
PAR.PROTECT
PROTECTION OFF
ENT
Press the
ENT
key once to display “PAR.PROTECTION”.
PAR.PROTECT

COMPLETE

INPUT ID NO.

ENT
Press the
ENT
key once to indicate "0000" and blink the cursor.
Note) If ID No. is "0000" (as factory set), press the
ENT
key to release
the protection.
INPUT ID NO.
0000
Enter ID No. “2234” by the key or the key.
INPUT ID NO.
2234
ENT
Press the
ENT
key once.
* If ID No. does not coincide, "INPUT ERROR!" appears, and the input
screen is resumed.
INPUT ID NO.

COMPLETE

――― Protection canceled. ―――
PAR.PROTECT
PROTECTION OFF
About the change of parameter setting
When you change parameters of converter in current use which analog output or alarm has been set, if you
change items which affect to the output or alarm, the output may change suddenly after display of
"**COMPLETE**" and may generate alarm. If, especially, the output signal is being used for control, perform the
signal lock on the system side prior to changing parameters.
Caution on change of parameter setting
When you change parameter settings, parameters will be saved in non-volatile memory on return to measuring
display. Saved parameters have been maintained even power is off. However, when you change the parameter
and turn off the power before returning to the measuring display, parameters will not be saved. Thus you should
set the parameter again.
CAUTION

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fuji Electric Time Delta-C FSV-2 and is the answer not in the manual?

Fuji Electric Time Delta-C FSV-2 Specifications

General IconGeneral
BrandFuji Electric
ModelTime Delta-C FSV-2
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals