4.10.7. Setting of the operation mode 
Description 
 Used to switch computation cycle and output cycle. 
 Settable range 
NORMAL  : Standard mode (factory-set value), computation/output cycle is approximately 0.5 seconds. 
HIGH SPEED : High speed response mode, computation/output cycle is approximately 0.2 seconds. 
 
For actual keying, refer to the typical operation indicated below. Set the protection to OFF beforehand. (See Section 4.4.1.) 
 
Operation 
(example) 
Switch the operation mode to the high speed response mode. 
Key operation  Description  Display 
   
 
▼
 
Press the key twice to display “OUTPUT SETUP”. 
OUTPUT SETUP 
 
 
ENT
 
 
▼
 
Press the
ENT
 
key once to display “ZERO ADJUSTMENT”. 
ZERO ADJUSTMENT 
SET ZERO
 
 
▼
 
Press the key for 10 imes to display “OPERATION MODE”. 
OPERATION MODE 
NORMAL
 
ENT
 
 
▼
 
Press the
ENT
 
key once to blink the cursor. 
OPERATION MODE 
NORMAL
 
 
▼
 
Press the key for 6 times to move the cursor. 
OPERATION MODE 
HIGH SPEED
 
ENT
 
 
▼
 
▼
 
▼
 
▼
 
Press the
ENT
 
key once to register. 
OPERATION MODE 
 COMPLETE 
 
↓ 
―――    “OPERATION MODE” has been registered.    ――― 
OPERATION MODE 
HIGH SPEED
 
ESC
 
Press the
ESC
key once and then press the key for 3 times to enter 
the measurement mode. 
        0.000      % 
        0.000      m3 
 
 
Reference 
The difference between standard mode and high speed mode 
High speed mode is unfit for the measurement when foreign objects or air bubbles are contained. 
Standard mode is about 10 times more resistant to entry of foreign objects or air bubbles than high speed mode.