4.10.7. Setting of the operation mode
Description
Used to switch computation cycle and output cycle.
Settable range
NORMAL : Standard mode (factory-set value), computation/output cycle is approximately 0.5 seconds.
HIGH SPEED : High speed response mode, computation/output cycle is approximately 0.2 seconds.
For actual keying, refer to the typical operation indicated below. Set the protection to OFF beforehand. (See Section 4.4.1.)
Operation
(example)
Switch the operation mode to the high speed response mode.
Key operation Description Display
Press the key twice to display “OUTPUT SETUP”.
OUTPUT SETUP
ENT
Press the
ENT
key once to display “ZERO ADJUSTMENT”.
ZERO ADJUSTMENT
SET ZERO
Press the key for 10 imes to display “OPERATION MODE”.
OPERATION MODE
NORMAL
ENT
Press the
ENT
key once to blink the cursor.
OPERATION MODE
NORMAL
Press the key for 6 times to move the cursor.
OPERATION MODE
HIGH SPEED
ENT
Press the
ENT
key once to register.
OPERATION MODE

COMPLETE

――― “OPERATION MODE” has been registered. ―――
OPERATION MODE
HIGH SPEED
ESC
Press the
ESC
key once and then press the key for 3 times to enter
the measurement mode.
0.000 %
0.000 m3
Reference
The difference between standard mode and high speed mode
High speed mode is unfit for the measurement when foreign objects or air bubbles are contained.
Standard mode is about 10 times more resistant to entry of foreign objects or air bubbles than high speed mode.