4.9.4. Setting the LCD indication
Description
Flow velocity indication
Selectable flow velocity units: m/s (if SYSTEM UNIT was set to METRIC) (See 4.8.1)
<Note> The decimal point position is fixed. (Decimal point 3 digits)
Flow rate indication
Selectable flow rate indications: Actual value reading, % reading.
<Note> The indication unit is as selected by FLOW UNIT. (See 4.8.2.)
Total indication
Selectable total indications: Actual total value reading (forward/reverse flow), total pulse count (forward/reverse flow).
<Note> The indication unit is as selected by TOTAL UNIT. (See 4.8.3.)
How to validate the indication
Set the DISPLAY setting mode to 1st ROW (for indication on 1st line) or 2nd ROW (for indication on 2nd line), and further select
indication contents.
For actual keying, refer to the typical operation indicated below. Set the protection to OFF beforehand. (See Section 4.4.1.)
Operation
(example)
Display the 1st line of LCD indication in percentages (%).
Key operation Description Display
Press the key twice to display “OUTPUT SETUP”.
OUTPUT SETUP
ENT
Press the
ENT
key once to display “ZERO ADJUSTMENT”.
ZERO ADJUSTMENT
SET ZERO
Press the key for 3 times to display “DISPLAY”.
DISPLAY
DISPLAY 1
ENT
Press the
ENT
key once to blink the cursor.
DISPLAY
DISPLAY 1
ENT
Press the
ENT
key again, and select “1ST LOW”.
1ST LOW
VELOCITY
Press the key twice to display “FLOW RATE(%)”.
1ST LOW
FLOW RATE(%)
ENT
Press the
ENT
key once, and select and fix “FLOW RATE(%)” to display
“1:DECIMAL POINT”.
1:DECIMAL POINT

.

Press the key once to shift the decimal point position to next place.
1:DECIMAL POINT

.

ENT
Press the
ENT
key once to register.
1:DECIMAL POINT

COMPLETE

――― FLOW RATE(%) indication has been set. ―――
1:DECIMAL POINT

.

ESC
Press the
ESC
key twice and then press the key for 3 times to enter
the measurement mode.
0.00 %
0.000 m3