-21-
4.3. Parameter initial value list
Factory-set value is shown below. (When parameter setting is not provided.)
Setting unit Setting range Initial value Setting value
1 Parameter protection No. of menu: 2 PROTECTION ON PROTECTION ON, PROTECTION OFF
2 ID No 0000 to 9999 0000 ID No. is invalid when 0000 is selected.
3 Language No. of menu: 5 English *1 English, Japanese, German, French and
Spanish
4
Measuring condition
System unit No. of menu: 2 Metric Metric or inch
5 Flow unit No. of menu: 18 m
3
/h L/s, L/min, L/h, L/d, kL/d, ML/d, m
3
/s, m
3
/min,
m
3
/h, m
3
/d, km
3
/d, Mm
3
/d, BBL/s, BBL/min,
BBL/h, BBL/d, kBBL/d, MBBL/d
6 Total unit No. of menu: 8 m
3
mL, L, m
3
, km
3
, Mm
3
, mBBL, BBL, kBBL
7 Outer diameter 6.00 to 6200.00mm 60.00mm [mm, in]
8 Pipe material No. of menu: 13
Sound velocity: 1000 to
3700m/s
PVC Carbon steel, Stainless, PVC, Copper,Cast
iron, Aluminum, FRP, Ductile iron, PEEK,
PVDF, Acrylic and PP
Pipe sound velocity
(Sound velocity: [m/s, ft/s])
9 Wall thickness 0.10 to 100.00mm 4.00mm [mm, in]
10 Lining material No. of menu: 8
Sound velocity: 1000 to
3700m/s
No lining No lining, Tar epoxy, Mortar, Rubber, Teflon,
Pyrex glass, PVC
Lining S.V. (Sound velocity: [m/s, ft/s])
11 Lining thickness 0.01 to 100.00mm – [mm, in]
12 Kind of fluid No. of menu: 18
Sound velocity: 300 to
2500m/s
Water Seawater, dist. water, ammonia, alcohol,
benzene, bromide, ethanol, glycol, kerosene,
milk, methanol, toluol, lube oil, fuel oil, petrol
and refrigerant R410
Fluid S.V. (Sound velocity: [m/s, ft/s])
13 Dynamic viscosity
coefficient
0.001 to 999.999
×10
-6
m
2
/s
1.0038
×10
-6
m
2
/s
[×10
-6
m
2
/s, ft
2
/s]
14 Sensor mounting method No. of menu: 2 V method V method, Z method
15 Sensor type No. of menu: 10 FSSA/FSSG FSSA/FSSG, FLS_12,/FLS_22,FSSC,
FSG_32, FSG_31/FSG_41,
FSSE/FSG_50,FSSF/FSG_51, FSD12, FSSD/
FSD22,FSSH/FSD32
16
Output condition
Zero adjustment No. of menu: 2 Clear (unadjusted) Clear, adjustment (Clear has been factory-set.)
17 Damping 0.0 to 100.0sec 5.0sec sec
18 Low flow cut 0 to 5m/s in terms of
flow velocity
0.150m
3
/h [(5) unit]
19
Display
Content of display 1st
line
No. of menu: 7 Flow velocity (m/s) Flow velocity, Flow rate, Flow rate (%), +Total
(Actual), +Total pulse, -Total (Actual) and -Total
pulse
20 Decimal point position
of display 1st line
.
(Fill in the specified digit)
21 Content of display 2nd
line
No. of menu: 7 Flow rate (m/s) Flow velocity, Flow rate, Flow rate (%), +Total
(Actual), +Total pulse, -Total (Actual) and -Total
pulse
22 Decimal point position
of display 2nd line
.
(Fill in the specified digit)
23
Analog output
Kind No. of menu: 2 Flow rate Flow verocity, Flow rate
24 Range type No. of menu: 4 Single range Single range, Auto 2 range, Bi-dir range and
Bi-dir Auto 2 range
25 Full scale 1 0, ±0.3 to ±32m/s in
terms of flow velocity
15.000m
3
/h [(5) unit]
26 Full scale 2 0, ±0.3 to ±32m/s in
terms of flow velocity
0.000m
3
/h [(5) unit]
27 Hysteresis 0.00 to 20.00 10.00% %
28 Burnout (current) No. of menu: 5 Hold Not used, Hold, Lower, Upper and Zero
29 Burnout timer 10 to 900sec 10sec sec
30 Output limit low -20 to 0% -20% %
31 Output limit high 100 to 120% 120% %
32 Rate limit 0 to 5m/s in terms of
flow velocity
0.000m
3
/h [(5) unit]
33 Rate limit timer 0 to 900sec 0sec sec
34
Total output
Total mode No. of menu: 3 Stop Start, Stop and Reset
35 Total rate 0.000000 to 99999999 0m
3
[(6) unit]
36 Total preset 0.000000 to 99999999 0m
3
[(6) unit]
37 Pulse width No. of menu: 7 50.0msec 5.0msec, 10.0msec, 50.0msec, 100.0msec,
200.0msec, 500.0sec,1000msec.
38 Burnout (total) No. of menu: 2 Hold Not used, hold
39 Burnout timer 10 to 900sec 10sec sec